Frequently Asked Questions

Troubleshooting Delmhorst Products

Troubleshooting Delmhorst Products

If the meter has a digital display, remove the battery and replace it with a new one. If applicable, reset the meter per the instructions in the owner’s manual, or reset it to default settings. If this does not resolve the issue, return the meter for repair. If the meter has an analog display, return the meter for repair.

... Permalink

Replace the battery with a new one. If the meter still reads out of tolerance, return it for repair.

... Permalink

This is usually an indication that the moisture content is less than the nominal low end of the meter's range. It can also be an indication of an 'open circuit' in the meter or electrode/prod.

... Permalink

Check to ensure the electrode pins and retainers are tight. Check the electrode cable connections at both ends. The cable may need to be reterminated or replaced.

... Permalink

If the reading is HI or over-range, this is usually an indication of a 'short' in the electrode, and it should be repaired. Electrical leakage in the electrode may also be the cause. Clean the electrode connector and the pins/retainers/insulation with denatured alcohol or a mild cleaner like Simple Green. If this does not resolve the issue, replace the electrode.

... Permalink

The original Delmhorst calibration was developed with a 4-pin electrode - four non-insulated steel pins, with 5/16in penetration in samples with uniform moisture distribution. Insulated pins read lower than non-insulated ones, so for best accuracy, a small correction is necessary. The default calibration for Delmhorst meters is for non-insulated pins, unless specified otherwise. Most meters are programmable with calibration for both insulated and non-insulated electrodes.

... Permalink
 