Frequently Asked Questions
Most FAQ
There is no set calibration schedule for our moisture meters. If the meter is used regularly, we recommend changing the battery at least every 3-6 months as this will help keep calibration indefinitely. Multi-function meters (pin/pin less/RH/T) may require more frequent battery replacement if the SCAN mode is used regularly.
Meter storage and usage temperatures may also affect battery life and so, calibration. Extreme temps in either direction will shorten battery life. We recommend removing the battery if the meter will not be used for 30 days or more. The user can check the calibration in the field by using an appropriate MCS (Moisture Content Standard) that we have available. Many meters have built-in calibration checks, as well.
Thermo-hygrometers can be returned to us for calibration checks and repair as needed or checked in the field using available salts solutions or a known reference sensor. If field testing of the sensor shows it to be out of calibration, the sensor should be replaced with a new one.
... PermalinkIf the meter has a digital display, remove the battery and replace it with a new one. If applicable, reset the meter per the instructions in the owner’s manual, or reset it to default settings. If this does not resolve the issue, return the meter for repair. If the meter has an analog display, return the meter for repair.
... PermalinkReplace the battery with a new one. If the meter still reads out of tolerance, return it for repair.
... PermalinkThis is usually an indication that the moisture content is less than the nominal low end of the meter's range. It can also be an indication of an 'open circuit' in the meter or electrode/prod.
... PermalinkCheck to ensure the electrode pins and retainers are tight. Check the electrode cable connections at both ends. The cable may need to be reterminated or replaced.
... PermalinkIf the reading is HI or over-range, this is usually an indication of a 'short' in the electrode, and it should be repaired. Electrical leakage in the electrode may also be the cause. Clean the electrode connector and the pins/retainers/insulation with denatured alcohol or a mild cleaner like Simple Green. If this does not resolve the issue, replace the electrode.